Header menu link for other important links
X
Development and validation of an HPTLC method for simultaneous estimation of thiocolchicoside and aceclofenac in combined dosage form
Pratima Syal, Madhusmita Sahoo, Rahul Raut, Asawaree Hable, Amruta Battewar, ,
Published in
2012
Volume: 25
   
Issue: 2
Pages: 133 - 137
Abstract
A simple, precise, rapid and accurate high-performance thin-layer chromatographic (HPTLC) method has been developed for the simultaneous estimation of thiocolchicoside (THIO) and aceclofenac (ACE) in bulk and pharmaceutical dosage form on silica gel precoated aluminum 60F 254 plates (10 cm × 10 cm) with 250 mm thickness. The separation was carried out in methanol-chloroform-water 9.6:0.2:0.2 (v/v/v) as mobile phase. The densitometric scanning was carried out at 254 nm. The R F values were found to be 0.70 ± 0.05 for THIO and 0.83 ± 0.05 for ACE. The linearity was obtained in the range 30-180 and 750-4500 ng per band for THIO and ACE with correlation coefficients r 2 = 0.9995 and 0.9993, respectively. The method was validated for specificity, linearity, accuracy, precision, limit of detection (LOD), limit of quantification (LOQ) and robustness. The percentage recovery for both the analytes was in the range of 98.7-101.2%. The proposed method was optimized and validated as per the International Conference on Harmonization (ICH) guidelines. {\textcopyright} Akad{\'{e}}miai Kiad{\'{o}}, Budapest.
About the journal
JournalJournal of Planar Chromatography - Modern TLC
ISSN09334173
Open AccessNo