Objectives: Olmesartan medoxomil (OLM) and metoprolol succinate (MPS) in fixed-dose combination (FDC) tablet formulation prescribed extensively. Stability indicating (SI) method for impurities and related substance (RS) test quantitates the amount of these analytes in formulation; the manuscript presents SI/RS-ultra-high performance liquid chromatography-photodiode array (UHPLC-PDA) method for OLM and MPS and their impurities. Materials and Methods: Well-resolved separation of all analytes was achieved with gradient elution on a Shimadzu on Shimpack GIST-C18 (100 mm x 2.1 mm, 2 µm) column maintained at 25°C. Mobile phase-A consist of 0.1% orthophosphoric acid in water and mobile phase-B was acetonitrile at a flow rate of 0.4 mL/min, data integrated at 225 nm and 16 min of short runtime for satisfactory elution of all peaks. Results: The proposed SI/RS-UHPLC-PDA method was developed and validated as per International Conference on Harmonisation (ICH) of Technical Requirements guidelines. The system suitability test complied by all eluted peaks of the interest with acceptable linearity, recovery, and precision. Specificity, robustness, and method sensitivity parameters were determined; all the parameters were found to be within the limits. All the impurities and stress-degraded peaks were well resolved. Conclusion: The proposed method was found to be simple, fast, linear, and accurate. Further, the method is precise, robust, and specific; suitable for routine IPQC during active pharmaceutical ingredient manufacturing, stability and impurity profiling studies of the titled bulk analytes. Furthermore, the method can be extended to assess the levels of impurities formed during life cycle of new FDCs of titled analytes. © 2023, Turkish Pharmacists Association. All rights reserved.